Specifications
English
11 – 2 Bluetooth specifications
Frequency band 2�4GHz ISM band (2�402-2�480GHz)
Transmitting power 2�5mW[dBm]
Modulation GFSK
Channels 40channel 2MHz spacing
Compatibility EN 300 328, EN 300 489-1, EN 301 489-17, EN 62479:2010
11 – 3 Classification of equipment
Type of protection against electric shock:
Class l equipment
Degree of protection against electric shock:
Type B applied part
(Applied part: handpiece)
Method of maintenance (cleaning) and sterilization recommended by the manufacturer:
See “6 Post-use Maintenance”
Degree of protection against ingress of water as detailed in IEC 60529 (EN 60529):
Foot control: IPX8 (Protected against the effects of continuous immersion in water)
Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide:
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide�
Mode of operation:
Non-continuous operation
11 – 4 Operation principle
Power is supplied to the control unit by operations on the foot control�
This makes the motor run and the bur attached to the handpiece rotate�
11 – 5 Warranty
NSK products are warranted against manufacturing errors and defects in materials� NSK reserves the right to analyze
and determine the cause of any problem� Warranty is voided should the product be not used correctly or for the
intended purpose or has been tampered with by unqualified personnel or has had non NSK parts installed� Replacement
parts are available for seven years beyond discontinuation of the model� Contact an authorized NSK dealer if repairs are
necessary�
11 – 6 Disposing of products
In order to avoid the health risks to operators handling the disposal of medical equipment, as well as the risks of
environmental contamination caused thereof, a surgeon or a dentist must to confirm the equipment is sterile�
Askspecialist firms who are licensed to dispose of specially controlled medical wastes to dispose the product for you�

Table of Contents

Other manuals for NSK Surgic Pro2

Related product manuals