Manual lubrication
Fig. 6
PANA SPRAY Plus
Nozzle
Caution
After cleaning the handpiece apply NSK PANA SPRAY Plus BEFORE autoclaving (Fig. 6). Make sure the CORRECT
application nozzle is connected to the plastic valve at the top of the spray can. Firmly insert the Nozzle into the rear of
the handpiece and activate the spray for 2-3 seconds until the spray exhausts from the handpiece head.
Always hold the NSK PANA SPRAY Plus can upright.
The spray lubricant is delivered from the can, into the handpiece, under
pressure. To prevent the handpiece from slipping from the Nozzle, always
hold the handpiece securily to the Nozzle.
Mount the arrow-head spray nozzle tip into the spray port and lubricate
chuck directly thorough the hole for bur attachment once a week.