Sterilization
Autoclave procedure
Use a brush to scrub debris from the handpiece. DO NOT use a wire
brush. Wipe clean with an alcohol-immersed cotton swab or cloth.
Lubricate the handpiece as previously described.
(Refer to Lubrication)
Insert the handpiece into an autoclave pouch and seal the pouch.
Autoclavable up to a max. 135°C
*
EN13060 4.6.3 recommends autoclaving for 3 minutes (minimum
holding time) at 134°C or 15 minutes (minimum holding time) at
121°C. NSK recommends Class B or S sterilization.
Sterilization by steam generation
autoclave is recommended.
Sterilization is required first time
you use and after each patient.