16
12-1 Specifications
<Control Unit> <Motor Handpiece>
Product Name VOLVERE i7 CONTROL UNIT Model VOLVERE i7 RMS
Model NE315 Speed 1,000min
-1
- 35,000min
-1
Rated input AC 100 V to AC 240 V, 50/60 Hz, 60 VA Dimensions 145 mm (L) × ø24.5 mm
Rated output DC 35 V 0.5 A Cord length 1.2 m
Dimensions 69 mm (W)×185 mm (D)×167 mm (H) Weight 245 g
Weight 930 g Remarks Combination of UHR45C and
GX35RM-B & SCD
<E-type Motor>
Model GX35EM-B & SCD
Speed 1,000min
-1
- 35,000min
-1
Dimensions 109.3 mm (L) × ø24.5 mm
Cord Length 1.2 m
Weight 170 g
<Foot Control>
Model FC-76
Dimensions 80 mm (W) × 136 mm (D) × 60 mm (H)
Cord length 2 m
Weight 221 g
Temperature Humidity Pressure
Operating conditions 0 - 40 °C 30 - 75% RH -
Transport/storage conditions -10 - 50 °C 10 - 85% RH 500 hPa - 1060 hPa
12
Specifications
<Handpiece>
Problem Cause Solution
The handpiece does not run while the
bur lock ring is in the LOCK position.
Entry of foreign matter in the ball
bearings or seizure.
Contact your Authorized NSK Dealer.*
Heat is generated during rotation.
Entry of foreign matter in the ball
bearings, causing wear of the
bearings.
Contact your Authorized NSK Dealer.*
Vibration or noise occurs during
rotation.
Entry of foreign matter in the ball
bearings, causing wear of the
bearings.
Contact your Authorized NSK Dealer.*
A bent bur is used. Replace the bur.
Run out of the bur is heavy.
Dust may be stuck in the chuck. Clean the inside of the chuck.
The chuck is worn. Replace the chuck.
The ball bearings are worn. Contact your Authorized NSK Dealer.*
The bur comes out. The chuck is loose. Tighten the chuck securely.
*Cannot be fixed at dental clinics, etc.