EasyManua.ls Logo

Ntron OxyTX - Page 13

Default Icon
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ntron OxyTX
Operation and Installation Manual
Revised:10/2013 ECN00008
Part No.:01-507
Ntron Ltd, Mullaghboy Industrial Park, Navan, Co Meath, IRELAND
Tel: + 353 (46)9071333 • Fax: + 353 (46)9071331 •
Email: info@ntron.com • Website: www.ntron.com/ntron
Ntron Ltd, Dallam Court, Dallam Lane, Warrington WA2 7LT, UNIITED KINGDOM
Tel: + 44 (161) 930 8690 • Fax: + 44 (161) 930 8691
View Engineer/Diagnostics Data (Menu Mode C:9):
This feature is a view-only feature. No configuration changes are possible from
within this menu.
This information assists in trouble shooting fault conditions present on the OxyTx.
Select menu mode C: 9 (refer to section 3) and press ENTER.
The display will alternate between the current value and code C: 9x: Where is
x is: 0 Sensor reading.
1 Calibration temperature, 8C.
2 Temperature compensation value 8C.
3 Sensor temperature, 8C.
4 Detector AtoD counts.
5 Reference AtoD counts.
The mode of operation can be selected by pressing the UP button.
Press MENU to return the instrument to its standard mode of operation.
Cross Reference (Menu Mode C:12):
No customer adjustment of this value is necessary. The set value of “1.00” should not
be changed.
Positive Zero Suppression (Menu Mode C:19):
No customer adjustment of this value is necessary. The set value of “0” should not be
changed.
Negative Zero Suppression (Menu Mode C:20):
No customer adjustment of this value is necessary. The set value of “0” should not be
changed.
(Menu Mode C:30):
No customer adjustment of this value is necessary. The set value of “OFF” should
not be changed.
P
AGE 13 OF 19

Related product manuals