CabinetSize 400 600
CabinetDimensions
60x28x46‐3/8
(1.52x.711x1.18m)
60x28x70‐3/8
(1.52x.711x1.8m)
CabinetVolume
45.1cu.ft.
(1.28cu.m)
68.4cu.ft.
(1.94cu.m)
NOTE:Theoutletsintheworkareaareenergizedaslongasthecabinetispluggedinandswitchedonthefront
panel.Unplugthecabinetbeforedecontaminationequipmentisplugged intotheseoutletsorrunthe
decontaminationpowercordsunderthefrontsealarea.
6. Useducttapeandplastictosealthefrontandexhaustarea.
BESURECABINETISTOTALLYSEALEDTOPREVENTANY
LABORATORYEXPOSURETODECONTAMINATIONGAS.
7. PerformdecontaminationprocedureperNSF/ANSI49,AnnexG.
Ifthecabinethasbeenusedtopreparehazardousdrugs,(chemotherapy),orothertoxicchemicals,decontaminationofthe
cabinetcannotbeaccomplishedbytheaboveprocedure.(Seesection6.6forguidelines)
PleaseconsultwithNuAire,Inc.aboutanyuniquecontaminationproblems.
Normally,nopreventivemaintenanceisrequiredontheinteriorofthecabinet(i.e.,theareabehindtheaccesspanel
containingtheHEPAfiltersandmotor(blowerassembly).Allrequiredadjustmentsinordertomaintainpropercabinet
airflowsareexternaltothecabinetinterior.Themotorislubricatedforlifeand isthermallyprotectedwithautomatic
reset.
7.2 FluorescentLampBulbReplacement
Thetwo(T8)fluorescentbulbsarecoolwhite,rapidstartandplacedexternaltothecabinettoaidmaintenanceand
minimizeheatbuild‐upwithinthecabinet.Theliferatingofthebulbis9000hoursbasedonthree‐hourburningcycles.
Toreplaceabulb,itisnecessarytoremovethelampassembly.
1. SwitchCabinetLightSwitchoff.
2. RemovethescrewsateachuppersideoftheControlCenterandallowtheControlCentertorotatedown,resting
onthesafetystraps.
3. Thebulbisnowdirectlyexposedforreplacement.
4. Thebulbisremovedbydisplacingthebulbtoonesideagainstthecompressiblebulbholderandliftingoutthe
bulb.
5. Reversetheproceduretoreinstallthelampassemblybeingcarefulnottopinchthesafetystraps,cableortubing
duringclosureofthecontrolcenter.