EasyManuals Logo

NuAire Labgard NU-430-400 User Manual

NuAire Labgard NU-430-400
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #87 background imageLoading...
Page #87 background image
OM0228 Page87of99
February/2014
ReadInputs
Readinputsallaservicetechniciantoexerciseorchecktheseinputfunctions.Theinputsmaybecheckedby
alteringthestateoftheinputfunction(i.e.slidingwindowposition)andmonitoringthechangeonthedisplay.













8.3 AirflowSensorPerformance
Verification
Theindividualairflowsensorscanberoutinelycheckedduringcalibration orindiagnosticstoassuretheyarereadingand
activewithintherangeofuse(0to200fpm)(0to1.02mps).Theairflowsensorscanalsobecheckedintherunmode
throughperformanceverification,forresponsivenesstochangingairflowconditions.
8.3.1 RunMode
Tochecktheairflowsensorinrunmode,firstallowthecabinettooperatenormallyforaminimumof5minutes.
Then,placearolledpieceofpaperoverthedownflowsensorintheworkzoneand leavethepaperonthesensor
foratleast2minutesandthenremove.Thisactionwillcausethecabinettogointoadownflowalarmcondition.
Theexhaustairflowreadingshouldincreaseduringthistest.However,thedownflowreadingshouldgodown
below25fpm(.13m/
s).Thereshouldalsobeanoticeableincreaseinmotor/blowernoise.Itwouldalsobe
recommendedtomonitormotor/
blowervoltageduringthetest.Themotor/blowervoltageshouldbemonitored
fromwhenthecabinetisrunningnormally.Duringthetest,whenthedownflowsensoriscovered,the
motor/blowervoltageshouldbesteadilyincreasing toslightlyunderlinevoltage.Whenthedownflowsensoris
uncovered,themotor/
blowervoltageshoulddecreaseandairflowreadingsshouldbewithinthecalibrationrange.
Ifthemotor/blowervoltagedoesnotchange,anairflowsensorproblemcouldexist.
PleaseconsultwithNuAireTechnicalService.
MENU
xx:xx
AM
xx/xx/xx
UV
FL
KEYPAD X = 0, Y = 0
DOWNFLOW SENSOR 60 FPM
EXHAUST SENSOR 1100 CFM
REMOTE OVERRIDE OPEN
NIGHT SETBACK OPEN
WINDOW POSITION NORMAL
PRESSURE SENSOR ACTIVE
AC LINE HZ 60 HZ
AC LINE VOLTS 115 VAC
MOTOR RPM 1000 RPM
NEXT
MEN
U
xx:xx
AM
xx/xx/xx
UV
FL
WINDOW HIGH LIMIT CLEAR
WINDOW HIGH CLEAR
WINDOW CLOSED CLEAR
WINDOW SAFE TRIPPED
NEXT

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NuAire Labgard NU-430-400 and is the answer not in the manual?

NuAire Labgard NU-430-400 Specifications

General IconGeneral
BrandNuAire
ModelLabgard NU-430-400
CategoryAir Conditioner
LanguageEnglish

Related product manuals