Step4:Verifytheoperationofthemanualmodepressureswitchatthispoint.Lowertheexhaust
volumeto100CFM(170CMH)lowerthanthestandardexhaustloweralarmlimit.
Turnoffmotor/blower
Blockoffsupplyductconnectionusingcardboard.
Lowerexhaustvolumetothepressureswitchlowtrippointstatedbelow.
Model LowAlarmLimit PressureSwitchLowTripPoint
CFM(CMH) CFM(CMH)
NU430400 641(1089) 541(919)
NU430600 935(1589) 835(1419)
Ifpresent,removecardboardfromsupplyairintakeandturnONtheunit’sblowerifitisnotONnow.
Remove1screwfromeachofthetoprightandleftsidesofthecontrolboxandrotatecontrolbox
forwardtoaccessandremovethepressureswitchadjustmentcap.Bringcontrolboxbacktovertical
positionandlooselythreadtheleftscrewofthecontrolbox.Pressureswitchmustbeadjustedin
verticalpositionsoleavetherightscrewofthecontrolboxoutsothepressureswitchcanbeaccessed
duringtheadjustment.Adjustasneeded(theadjustmentscrew),untiltheinternalsupplyblower/motor
isdeactivated.Thenraisetheexhaustvolumebackuptothestandardloweralarmlimittoensurethe
internalsupplyblower/motorturnsbackON.Finally,adjusttheexhaustsystembacktothecorrect
inflowvelocityandexhaustvolume.Reattachcontrolboxandinstallpressureswitchcap.
Step5: Ifpresent,removedirectreadinginstrumentfromwindowaccessopenarea.
7.5.3.2 E
xhaustVolumeCalibrationwiththeNU951012MotorizedButterflyValve

Ifdesired,activateexhaustcontrol/autozero.
TheexhaustcontrolautozerooptionallowstheNU951automaticdampertocheckitselfonceevery24hours
fromitsoriginalcalibrationpoint.Duringtheairflowcalibrationprocedure,oncethecertifiercalibratesthe
exhaustvolumethecontrolsystemwillrememberthedamperpositionindegreesatthat point.Thenbased
ontheprogrammedtime,theexhaustsystemwillperformthefollowingautozerosequence.First,freezethe
exhaustdisplaytoitssetpoint.Thenmovethedamperbacktoitsoriginalcalibrationpoint.Allowthedamper
tocontrolnormallyfortwominutesthenunfreezethedisplay.Thissequencewillthen beperformedatthe
programmedtimeonceevery24hours.TheautozerosequencecanbecheckedbypressingTESTtoinitiatea
fourminutetestsequence.
Toactivatetheexhaustcontrol/autozero,performthefollowing:
SelectOptionSetUp.
MEN
xx:xx
AM
xx/xx/xx
SETPOINTS/LIMITS
SENSOR SETUP
NIGHT SETBACK
SERVICE
OPTION SETUP
DIAGNOSTICS
CABINET RESET
CABINET TYPE
DAMPER CALIBRATION
UV
FL
UP
DOWN
ENTER

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NuAire Labgard NU-430-400 and is the answer not in the manual?

NuAire Labgard NU-430-400 Specifications

General IconGeneral
BrandNuAire
ModelLabgard NU-430-400
CategoryAir Conditioner
LanguageEnglish

Related product manuals