OM0228 Page69of99
February/2014
8.2.3 CabinetType/MotorType
Thecabinettypecanbeverifiedinthecontrolsystemandisfactorysetandshouldn’trequirealteration.The
cabinettypedefaultinformationcontrolsunitofmeasure,setpointsandlimitsbasedonthetypeandsizeofcabinet.
Toverify,pressCABINETTYPE.Thecurrenttypeofcabinetwillbedesignated.Againtoverify
,pressthecorrect
TypeandtheModel/Sizeofthecurrentselectedcabinetmodelwillbedesignated.Onceverifiedthecorrecttype
andmodelaredesignated,thenpressBACKtoreturntoCalibration/Servicemenu.
PressSET
MOTORTYPEtoverifycorrectsetting.UponaMASTERRESET,themotortypeisdefaultedtoAC.The
NU‐430Series60requiresthemotortypeselectedtobeDC.Alwaysverifymotortypewhenverify
ingcabinet
type.
UV
FL
UV
FL
MEN
xx:xx AM xx/xx/xx
TYPE A2 (70% RECIRCULATE/ 30% EXHAUST)
TYPE B1 (30% RECIRCULATE/ 70%EXHAUST)
TYPE B2 (100% EXHAUST)
SET OEM DISPLAYED
SET MOTOR TYPE
UP
DOWN
ENTER
CURRENTLY B2 NU-430/435-400
MENU
xx:xx
AM
xx/xx/xx
UP
DOWN
SAVE
B2 NU-430/435-300
B2 NU-430/435-400
B2 NU-430/435-500
B2 NU-430/435-600
B2 NU-430/435-800
B2 NU-430-400(E)
B2 NU-430-600(E)
B2 NU-430-400(D)
B2 NU-430-600(D)
MORE
CURRENTLY B2 NU-430/435-400