6
2. TECHNICAL SPECIFICATIONS
2.1. TECHNICAL SPECIFICATIONS TABLE
TECHNICAL SPECIFICATIONS
KN 72
KN 120
KN 294
KN 504
Temperature Range
0.0°C - 10.0°C
Temperature Sensor
Pt-100
Temperature Setting And
Reading Sensitivity
0.1°C
Thermostat Sensitivity
± 0.1°C
Temperature Homogeneity
at 4°C
± 1.5°C
Temperature Control
Temperature measurement from a liquid similar to blood
Nominal Storage Capacity
(450 ml Blood Bags)*
72 pcs.
120 pcs.
294 pcs.
504 pcs.
Capacity (liter)
200
303
630
1090
Refrigerant
R134A
Alarm Back-Up
Automatically re-chargeable battery for 12-hour
Alarm System
Audible and visible
Temperature Alarm Range
± 2°C
Chamber Illumination
(Fluorescent)
15 Watt
18 Watt
30 Watt
2x30 Watt
Door
Insulated steel frame with triple glass window and key lock
Drawers
Stainless steel, drawers
Number Of Drawers
3 pcs.
5 pcs.
6 pcs.
12 pcs.
Control System
N-SmArt™ Programmable microprocessor control system
Temperature Indicator
Colorful LCD
Temperature Recorder
Battery operated 7-day circular chart with key lock
Internal Memory
10-year memory, one record per hour
USB
Standard
Ethernet
Standard
RS 232
Standard
Central Alarm Outlet
Standard
Central alarm system
Standard
Power Supply
230 V, 50/60 Hz
Power Consumption
330 Watt
330 Watt
800 watts
1000 Watts
Internal Material
Stainless steel
External Material
Epoxy polyester powder coated stainless steel
Internal Dimensions (WxDxH)
(mm)
480x545x770
480x545x1170
735x635x1350
1274x635x1350
External Dimensions(WxDxH)
(mm)
610x805x1255
610x805x1855
917x895x1985
1544x900x1985
*Capacity of the instrument depends on the loading method.