Kalix Sanitisation Procedure.
TECHNICAL
BULLETIN
OAS0033A
181218
OAS0033A
5
Disconnect the cooler from mains water and electrical supply. Flush the cold tank and drain
tube by following [STEP 6].
Sanitization Instructions Additional Parts:
10. Sanitizing solution
Prepare a 2000 ppm Peracetic acid solution by mixing 40ml of 5% Peracetic acid per liter of
room temperature water.
11. Additional Parts Sanitization
Diagram 12.1 Diagram 12.2
Unscrew spout from water outlet. (Diagram 12.1)
Remove blue tubing between water outlet and solenoids on cook and cold sides. (Diagram
12.2)
Remove water outlet from unit.
Undo cable ties holding solenoids and Tee-piece in position. Dispose of the blue tubing
connecting the solenoids and Tee-piece
Disconnect the 4 ft blue tubing from the Tee-piece and bulk head fitting. Retain tee-piece for
sanitization.
Disconnect sanitizing cartridge from unit. Remove bulk head fitting from the condenser on
unit and retain for sanitization.
Remove/Dispose blue tube between cooling tank and solenoids.