Calibration
Minimum Light Settings
Black Panel light level White Panel light level
0.70 to .085 Foot Lamberts 145 to 155 Foot Lamberts
7. Carefully (see Note of step 5) select maximum light settings by clicking on one of the maximum brightness or contrast sliders.
8. Using a light meter, verify that the black and white panels on the center of the monitor produce the following maximum light
levels:
Maximum Light Settings
Black Panel light level White Panel light level
1.1 to 1.25 Foot Lamberts 195 to 205 Foot Lamberts
9. Click on the Monitor pull down button select the Right monitor.
10. Repeat the Minimum and Maximum light setting adjustment for the right monitor
11. When both monitors produce the correct light levels select the File pull down menu and select the Save Workstation
Calibration File to save the new monitor brightness and contrast calibration data.
12. From the File pull down menu select the Download Workstation Calibration data to the 9800.
13. From the Tools pull down menu select Disconnect.
14. Turn the 9800 system off, wait for about 10 seconds then turn the system back on.
Using the Utility Suite CD
1. Connect the utility cable to the laptop and the back of the workstation.
2. Start the Utility Suite software.
74
Service
Periodic Maintenance
Contents
Schematics
Illustrated Parts
Installation