Calibration
14. To check the alignment of the right monitor Click on the Right tab on the form and repeat the steps for the left monitor
alignment.
15. If changes are made in monitor brightness and contrast settings, save the changes to the system by selecting the File pull
down menu, Save, then To System ,then Both. The message
YOU MUST REBOOT THE SYSTEM IN ORDER FOR THE
SAVED DATA TO BE APPLIED TO THE SYSTEM
will be displayed. Click on Ok.
16. Turn the 9800 system OFF, wait for about 10 seconds then turn the system back ON.
Light Source Verification / Calibration
The following equipment is required to calibrate and verify the light source.
Light meter
Fostec light source and head with mounting bracket.
Fostec light source control PCB
76