Replacement
Follow the steps below to replace the Generator Driver PCB
WARNING: Dangerous voltages are present when covers are removed from the C-Arm.
1. Place the Workstation power switch in the OFF position and disconnect Workstation's AC power plug from AC outlet.
2. Remove the Rear Cover and disconnect the steering linkage.
3. Remove seven screws that secure the Generator Driver Cover and remove the cover.
4. Disconnect the cabling from the Filament Driver PCB.
5. Remove two screws from the Filament Driver heat sink.
6. Remove the Filament Driver PCB.
7. Remove five screws that secure the Generator Driver PCB.
8. Remove the Generator Driver PCB.
High Voltage and Control Cable Assembly
WARNING! The High Voltage and Control Cable Assembly carries highly lethal DC voltage even when
the system is powered down and disconnected from the power line. Lethal static charges can
remain on the cable long after the system has been turned off. Do not attempt to service or
replace this cable until you are sure that the C-Arm is completely off, and you have taken
steps to neutralize all electrostatic charges.
The High Voltage and Control Cable is the thickest cable assembly in the C-Arm. It runs through a large strain relief and
grommet on the front of the C-Arm chassis to the C itself, where it services the X-ray tube, Collimator, Image Intensifier, and CCD