EasyManua.ls Logo

OEC 9800C-Arm - Page 430

OEC 9800C-Arm
490 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Replacement
High Voltage and Control Cable Assembly
The High Voltage and Control Cable (GE OEC Part Number 901141) is a complex wiring assembly that contains many
conductors, including the thick, highly insulated positive and negative cables that carry high voltage from the High Voltage tank to
the X-ray tube. Only the entire cable is field replaceable. It is not possible to replace only a wire or two in the cable Assembly.
You can, however, repair or replace broken wires or connectors at either end of the cable.
The following table shows the principal connections the High Voltage and Control Cable Assembly makes between the C-Arm
chassis and components in the C. Refer to the C-Arm Interconnect Diagram (878980) for more information such as voltage
levels and signal names.