Replacement
2. Position the C in a horizontal position with the Image Intensifier furthest from the C-Arm chassis.
3. Lock the C-Arm brake.
WARNING: Failure to install the C-Arm Safety Bolt may allow the gas spring to rotate the L-Arm
very quickly and very forcefully when you remove the Image Intensifier. This could
result in serious injury or equipment damage.
4. Remove the L-Arm end cap and install the C-Arm safety bolt as described in the L-Arm replacement procedure provided
elsewhere in this chapter
5. Remove the CCD Camera as described elsewhere in this chapter.
6. Remove four socket hex bolts that secure the Image Intensifier assembly to the C-arm weldment.
7. Lift the Image Intensifier assembly up and disconnect all wiring from it.
8. Remove the Image Intensifier.
9. Reassemble in reverse order, installing replacement Image Intensifier. Check Beam Alignment.