EasyManua.ls Logo

OHAUS Adventurer AX224 - User Manual

OHAUS Adventurer AX224
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Adventurer
TM
Balances
Instruction Manual

Table of Contents

Other manuals for OHAUS Adventurer AX224

Questions and Answers

  • M
    Miss Denise BrownAug 13, 2025
    What to do if my OHAUS Scales will not turn on?
    • M
      michellejohnsonAug 13, 2025
      If your OHAUS Scales are not turning on, the issue is likely a power problem. Verify the power connection and voltage to ensure the scales are receiving power.
  • P
    Philip KingAug 17, 2025
    Why can't I calibrate my OHAUS Scales?
    • J
      Jamie LaneAug 17, 2025
      Several factors can prevent calibration of your OHAUS Scales. Here are a few things to check: * Ensure the Calibration Menu lock is off. * Confirm that Approved Mode is off. * Make sure the environment is stable; relocate the scales if necessary. * Verify that you are using the correct calibration masses.
  • D
    danielwoodsAug 21, 2025
    What does 'Low Reference weight' mean on my OHAUS Adventurer AX224, and how to fix it?
    • L
      Lance DanielAug 21, 2025
      A 'Low Reference weight' message on your OHAUS Scales means the reference weight is too small. Increase the sample size to resolve this.
  • M
    Matthew WellsAug 25, 2025
    What does 'Invalid Piece Weight' mean on my OHAUS Adventurer AX224, and how to resolve it?
    • L
      Lynn WebbAug 25, 2025
      An 'Invalid Piece Weight' message on your OHAUS Scales indicates that the average piece weight is too small. To fix this, increase the average piece weight.
  • S
    Shawn WeberAug 28, 2025
    Why are my OHAUS Adventurer AX224 showing poor accuracy?
    • A
      Amanda SerranoAug 28, 2025
      Poor accuracy with your OHAUS Scales can stem from a couple of causes. First, the scales may need calibration. Try performing a calibration. Second, an unstable environment can affect accuracy. Relocate the scales to a more suitable, stable location.
  • C
    Carl TylerSep 1, 2025
    Why do my OHAUS Scales display 'Operation Timeout'?
    • F
      Faith FlynnSep 1, 2025
      An 'Operation Timeout' error on your OHAUS Scales indicates that the weight reading is not stable. Try moving the balance to a more suitable location to ensure stability.
  • A
    Angela GouldSep 5, 2025
    Why can't I change the menu settings on my OHAUS Adventurer AX224 Scales?
    • K
      Kristopher CarrSep 5, 2025
      If you are unable to change menu settings on your OHAUS Scales, it could be due to a locked sub-menu. Try unlocking the sub-menu. Also, ensure that Approved Mode is turned off.
  • J
    Jordan SantiagoSep 8, 2025
    What does '------' mean on OHAUS Adventurer AX224?
    • P
      Peter MillerSep 8, 2025
      The '------' display on OHAUS Scales indicates that the balance is busy. It may be performing a tare, zeroing, printing, or waiting for a stable weight. Wait until the process is completed.

Summary

Introduction and Safety

Description and Features

Overview of the Adventurer balance's capabilities, including touch controls and applications.

Safety Precautions

Essential safety warnings and precautions for installation, use, and handling of the balance.

Installation and Basic Setup

Unpacking and Location

Steps for safely unpacking the balance and selecting an appropriate location for optimal performance.

Leveling Equipment

Instructions on how to level the balance using the bubble indicator and leveling feet.

Initial Calibration

Procedures for performing initial calibration after installation or relocation to ensure accurate measurements.

Operation and User Interface

Display and Home Screen

Explanation of the balance's touch-sensitive display, controls, and the main application screen layout.

Principal Functions and Menu

Overview of the balance's main functions, buttons, and menu navigation structure.

Parts and Features

Identification and description of key physical parts for both draft shield and non-draft shield models.

Applications

Weighing

Guide on using the Weighing application to determine the weight of items in the selected unit.

Parts Counting

Instructions for using the Parts Counting application to count samples of uniform weight.

Density Determination

Detailed steps for measuring the density of solids and liquids using various methods.

Check Weighing

Instructions on how to use the Check Weighing application to compare sample weight against target limits.

Formulation

Guide for using the Formulation application for compounding and recipe making.

Menu Settings and Configuration

Menu Navigation

Overview of how to navigate through the balance's menu structure to access settings.

Calibration

Explanation of the different calibration methods available, including internal, automatic, and span calibration.

Balance Setup

Guide to customizing balance settings such as language, user preferences, and display options.

Weighing Units

How to activate and select desired units of measure for weighing operations.

Communication

Instructions for configuring external communication methods like RS-232 and USB.

Technical Data, Maintenance, and Printing

Printer Interface Setup

Steps for connecting, configuring, and testing the interface with printers or computers.

Maintenance

Guidelines for periodically verifying calibration and performing routine cleaning of the balance.

Specifications

Detailed technical specifications of the Adventurer balances, including capacity, readability, and dimensions.

USB Interface

Information on connecting the balance to a computer using the USB interface and installing virtual port software.

Compliance and Warranty

Compliance

Information regarding the balance's compliance with relevant EMC, Low Voltage, and weighing instrument directives.

Limited Warranty

Details of the product's warranty against defects in materials and workmanship.

OHAUS Adventurer AX224 Specifications

General IconGeneral
Capacity220 g
Pan Size90 mm
CalibrationExternal
Power SupplyAC Adapter (Included)
Operating Temperature Range10 °C to 40 °C
DisplayLCD
Repeatability (typical)0.1 mg
Stabilization Time2 seconds
Weighing Unitsg, lb, oz
CommunicationRS232
Readability0.1 mg