Calibration Guide Adventurer
Pro
Ohaus Corporation www.ohaus.com i Adventurer
Pro Calibration Guide
PREPARATION
Adventurer Pro balances have two calibration methods: Span and Linearity. Some models have
Internal Calibration (InCal
) for Span calibration.
Ensure that the scale is level, stable, and undisturbed during calibration.
(Otherwise the process could fail.)
Allow scale to warm up for five minutes after stabilizing to room temperature.
Use suitable calibration masses (available from Ohaus).
Exact weight value depends on model. Follow prompts for indicated values to use.
(Weight values here are examples only.)
Calibration may be locked out. To unlock it, use the Unlock menu. If the scale is set to
Legal for Trade operation, a seal must be broken to unlock it.
(See the Instruction Manual.)
Note: To cancel calibration, press:
INTERNAL CALIBRATION
(No external weight needed.)
5
(Release button)
1
1
When calibration finishes, balance returns to currently selected mode.
SPAN CALIBRATION
5
(Release button)
(If InCal appears press No.)
1
1
(Acquiring zero value)
1
(Place weight on pan.)
(If SPAN DONE flashes, remove mass from Pan.)
(Press No to change.)
Balance returns to weighing mode.

Other manuals for OHAUS Adventurer Pro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the OHAUS Adventurer Pro and is the answer not in the manual?

OHAUS Adventurer Pro Specifications

General IconGeneral
DisplayLCD
Power SupplyAC Adapter
CommunicationRS-232
Operating Temperature Range10°C to 40°C
Unitsg, oz, ozt, dwt, ct, gn, tl, tola
Pan SizeØ 90mm to 190 x 200mm (model dependent)
CalibrationInternal or external (model dependent)
Platform SizeØ 90mm to 190 x 200mm (model dependent)

Related product manuals