9. CONTINUOUS WAVE MODE CERTIFICATION (Average Power)
WATTS
SELECT
POWER
ON
OFF
DIFF.
WATTS
ALLOWABLE
RANGE
3.7 - 6.3
SAT.
UNSAT.
Average of 3 Readings
5
WATTS
SELECT
POWER
ON
OFF
DIFF.
WATTS
ALLOWABLE
RANGE
7.4 - 12.6
SAT.
UNSAT.
Average of 3 Readings
10
WATTS
SELECT
POWER
ON
OFF
DIFF.
WATTS
ALLOWABLE
RANGE
11.1 - 18.9
SAT.
UNSAT.
Average of 3 Readings
15
WATTS
SELECT
POWER
ON
OFF
DIFF.
WATTS
ALLOWABLE
RANGE
14.8 - 25.2
SAT. UNSAT.
Average of 3 Readings
20
WATTS
SELECT
POWER
ON
OFF
DIFF.
WATTS
ALLOWABLE
RANGE
SAT.
UNSAT.
Average of 3 Readings
WATTS
SELECT
POWER
ON
OFF
DIFF.
WATTS
ALLOWABLE
RANGE
SAT. UNSAT.
Average of 3 Readings
10. PULSED MODE CERTIFICATION (Amplitude Modulated Waveform)
MAX.
PULSE
MODE
SETTING
POWER
ON
OFF
DIFF.
WATTS
CALCULATIONS
(Pp)
Average of 3 Readings
= Measured Average Power (Av)
Pp
R
TPA
CALC.
AVERAGE
POWER
(Pp / R
TPA
)
Difference
Between
Measured AV
And
Calculated AV
Is Difference < ± 0.6% Of
(Pp / R
TPA
)
YES NO
REMARKS
11. SHORT TERM LIFE TEST COMPLETE? YES NO
12. ADDITIONAL TEST (Describe in Detail):