Chapter 3 Preparation and Inspection
67
EVIS LUCERA GIF/CF/PCF TYPE 260 Series OPERATION MANUAL
Inspection for smooth operation
1. Confirm that both the UP/DOWN and RIGHT/LEFT angulation locks move
all the way in the “F ” direction (the GIF-N260 has only the UP/DOWN
angulation lock).
2. Turn the UP/DOWN and RIGHT/LEFT angulation control knobs slowly in
each direction until they stop, and return them to their respective neutral
positions (the GIF-N260 has only the UP/DOWN angulation control knob).
Confirm that the bending section angulates smoothly and correctly, that
maximum angulation can be achieved, and that the bending section returns
to its neutral position.
3. When the UP/DOWN and RIGHT/LEFT angulation control knobs are turned
to their respective neutral positions as shown in Figure 3.7, confirm that the
bending section returns smoothly to an approximately straight condition
(except GIF-N260).
Figure 3.7
4. When the UP/DOWN angulation control knob is turned to its neutral position
as shown in Figure 3.8, confirm that the bending section returns smoothly to
an approximately straight condition (for GIF-N260 only).
Figure 3.8