SZ-CHI
A.
OUTLINE
OF
PRODUCT
(4)
Be
sure
to
attach
the
diopter
adjustment
ring
to
the
eyepiece
sleeve.
(5)
Use
TL2
transformer.
(6)
When
observing
a
specimen
with
low
reflectivity,
a
ghost
may
appear
in
the
visual
field
with
a
low
magnification.
(7)
When
room
temperature
(or
storing
temperature)
rises
over
50°C,
the
polarizing
performance
is
degraded
and
not
restored.
(8) The
quarter
wave
plate
compatibility
is
not
guaranteed.
Therefore,
do
not
change
the
combination
at
shipment.
(9) The
optical
axis
and
parfocality
need
adjustment
after
the
illuminator
is
attached
to
the
microscope.
If
they
~re
not
adjusted,
the
quality
will
be
as
follows:
Centration
of
left
and
right
tubes
0.3
max.
in
the
up/down/outward
direction
and
0.4
inward
Decentration
of
image
by
interpupillary
distance
adjustment
···
0.25mm
max.
on
image
surface
Decentration
of
image
when
zooming
•••
0.7mm
max.
on
image
surface
Parfocality
between
left
and
right
tubes
±2
diop
at
low
magnification,
±3
diop
at
high
magnification
A-2