Disclaimer: Information provided in this guide is not exhaustive. Please refer to the complete Instructions For Use for each
item of equipment which should be thoroughlyread and understoodprior to handling, use and storage of the equipment.
UHI-4 Insufflator Quick Reference Guide
Insufflator Controls Cavity Modes
Flow Rates
There are two cavity modes Normal and Small which regulate the pressure and flow rates.
The are three flow settings:
Note: The cavity mode cannot be changed once insufflation has started, if required stop
insufflation and then change mode.
Warning: when using a Veress needle always use the LOW flow mode as insufflation in other
modes is dangerous because the cavity pressure may exceed the set pressure.
Press to toggle between normal and small
Cavity mode Setting pressure range Note
NORMAL
3 - 25 mmHg
(0.4 - 3.3 kPa)
To increase the pressure from 20
to 21 mmHg, keep pressing “+”
for at least three seconds
SMALL
3 - 15 mmHg
(0.4 - 2.0 kPa)
Cavity mode Setting pressure range Note
NORMAL
HIGH
MED
LOW
20 - 45
1.1 - 1.9, 2 - 5
0.1 - 1.0
SMALL
HIGH
MED
LOW
6 - 10
1.1 - 1.9, 2 - 5
0.1 - 1.0
Press to toggle between flow rates
Smoke Evacuation
Connector
Power Switch
Set Pressure Flow Rate
Actual Pressure Volume Indicator
CO
2
Connector Smoke Evacuation
Speed
Start/Stop
Issue 1
05/02/2021
KMF-08929
NCDR 02867

Other manuals for Olympus UHI-4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus UHI-4 and is the answer not in the manual?

Olympus UHI-4 Specifications

General IconGeneral
BrandOlympus
ModelUHI-4
CategoryMedical Equipment
LanguageEnglish

Related product manuals