EN-6 Vanta — Getting Started Guide
Trademarks
SD, miniSD, and microSD logos are trademarks of SD-3C, LLC.
All brands are trademarks or registered trademarks of their respective owners and third party entities.
Radiation Profiles
The table below represents the upper bounds on the worst case (maximum power and minimum beam
filtration) using a 316 stainless steel target. Note that these beam conditions do not represent typical
use values or a combination of settings typically available from the factory. To convert from µSv/h to
mR/h, divide the value by 10.
Model
Beam
conditions
Trigger 5cm 10 cm 30 cm
VLW, Element 35 kV,
50 µA,
Al filter
BK
a
a. BK = background reading (< 1 µSv/h)
4 µSv/h 2 µSv/h BK
VCR, VCW 40 kV,
100 µA,
open filter
BK 25 µSv/h 10 µSv/h BK
VMR, VMW,
VCA,
Element-S
50 kV,
80 µA,
open filter
BK 4Sv/h 2Sv/h Sv/h