Section 7 - Specifications
Target Plate Temperature Range: Amb. + 11 to 215°C (Amb. + 20 to 420°F)
Accuracy: ±0.5°C, ±0.25% rdg
(±0.9°F, ±0.25% rdg)
Stability: ±0.1°C (±0.2 °F )
Ambient Environmental Conditions
Temperature: 5 to 45 °C (41 to 113°F)
Humidity: 0 to 90% RH, Non-condensing
Target Plate Size: 63.5 mm (2.5")
Target Plate Emissivity: 0.95*
Internal Control Sensor: Platinum RTD, Class A, Alpha = 0.00385
Reference Sensor: Platinum RTD, Class A, Alpha = 0.00385
Warm-up Time: See Figure 7, Section 3.5
Cool-down Time: See Figure 7, Section 3.5
Power
BB702: 115 VAC~, 50/60 Hz, 75W.
BB702-230VAC: 230 VAC~, 50/60 Hz, 75W
Dimensions: 152 x 305 x 280 mm
(6 H x 12 W x 11"L)
Weight: 7.2 kg (16 lb)
Installation Category II
* Reference to 8-14 micons wavelength bandwidth.