Omega CT485B-110V-G-AL - Temperature Calibration Procedure

Omega CT485B-110V-G-AL
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.4 Calibrating Temperature Readings
1. To calibrate temperature, set the instrument to °F (DO NOT USE °C) and place
the sensor probe in a known, stable environment between 60°F and 90°F. Read
the display after allowing the probe to reach equilibrium after a 15 minute
warm-up period.
2. If the reading is TOO LOW by n counts, DECREASE the temperature code (code
following the letter “T”) by n counts. If the reading is TOO HIGH by n counts,
INCREASE the temperature code by n counts.
Changing the temperature code will not affect the humidity calibration.