9.2.10  SP CNF (Setpoints 1 & 2 Configuration)
Refer to Section 10 for an in-depth discussion of these features.
Setpoint configuration is used to select:
• the active zone of each setpoint to above and below the setting
[SPC.1 & SPC.4]
• whether the open-collector output is on or off [SPC.2 & SPC.5]
• whether the reading compared with the setpoints is filtered or unfiltered 
[SPC.3 & SPC.6]
• enabled or disabled setpoints [SPC.7]
• enabled or disabled setpoint LED displays [SPC.8]
MENU MIN/MAX/MENU *
BUTTON BUTTON
MAIN MENU SUB MENU DESCRIPTION
“SP CNF”: SETPOINTS 1 & 2 CONFIGURATIONS
Setpoint 1 Active Zone:
“SPC.1=0
” Above
“SPC.1=1” Below
Setpoint 1 open-collector or relay output
Active ON or OFF:
“SPC.2=0
”On
“SPC.2=1” Off
Filtered/unfiltered reading compared 
with Setpoint 1 value:
“SPC.3=0” Unfiltered
“SPC.3=1” Filtered
Setpoint 2 Active zone:
“SPC.4=0” Above
“SPC.4=1” Below
Setpoint 2 open-collector or relay output
Active ON or OFF:
“SPC.5=0
”On
“SPC.5=1” Off
Filtered/unfiltered reading compared 
with Setpoint 2 value:
“SPC.6=0” Unfiltered
“SPC.6=1
” Filtered
Setpoints 1 & 2 action:
“SPC.7=0” Enabled
“SPC.7=1” Disabled
Setpoint 1 & 2 LEDs action:
“SPC.8=0” Enabled
“SPC.8=1” Disabled