Troubleshooting
7. Troubleshooting
7.1 Calibration
7.1.1 Reset Individual Channels
To reset the channel if you have trouble calibrating an individual analog input or output channel,
follow these instructions:
1.
Click the Diagnostics button in the Home
ribbon to open the Diagnostics tree in the
left panel.
Select System Resets in the left panel
and then select the Reset Calibration tab
in the screens to the right.
2.
In the Reset Individual Calibration
Constants section of the Reset
Calibration screen:
a. Select the Type radio button for
the channel.
b. Input the channel number in the
Number field.
c. Click Reset Calibration
Constants.
3.
At the prompt, click Yes to reset the
channel.