OMNI 4000/7000 Operations and Maintenance Guide Rev F
7-27
Maintenance
9.
Adjust the Signal Generator or configure
the Transmitter (whichever is connected)
to output the appropriate Low value
according to the selected calibration units.
10.
Confirm that the Live Input (mA/Ohms)
field value is close to the applied value
from the field transmitter or
signal generator.
If the value is not close to the applied
value, check:
The wiring is connected to the
correct channel.
The hardware jumper settings.
See the following sections in the Installation Guide for more information on wiring and
jumper settings:
Section 3.7.5 Process I/O Modules for tables on back panel wiring.
Appendix A: Jumper Settings for hardware jumper settings.
11.
Wait 20 to 30 seconds for the readings to
stabilize.
12.
While the Applied Input Value? field is
highlighted, press OK.
13.
Using the numeric keypad, enter a value
that exactly matches the value at the field
transmitter or the signal generator.
Press OK.
Low values are 4.00 mA for a
420 mA input, 1 V for a 15 V
input, and 25 ohms for an
RTD input.