Adjustable Settings
Automated Insulin Delivery
Target Glucose is the only adjustable setting that directly impacts
automated insulin delivery.
• Customizable to user needs
• 110, 120, 130, 140, 150 mg/dL
• Up to 8 time segments
Target Glucose is used for both the target for automated insulin delivery
in Automated Mode as well as the target for correction bolus dosing in
Automated and Manual Modes.
Manual Basal Delivery SmartBolus Calculator
Adjusting Basal Program segments only
impacts Manual Mode basal insulin delivery.
• Basal Program (Manual Mode)
SmartBolus Calculator settings impact suggested
bolus doses in Automated and Manual Modes and
include:
Target Glucose and Correct Above Values
Insulin to Carb Ratio
Correction Factor
• Duration of Insulin Action
• Reverse Correction
• Minimum Glucose for Calculations
The Activity Feature
The Activity feature is an optional feature in Automated Mode. When enabled,
the Activity feature reduces automated insulin delivery and temporarily sets
the Target Glucose to 150 mg/dL. The Activity feature is ideal during times of
anticipated decreases in insulin needs such as aerobic exercise, or other periods
of caution towards hypoglycemia.
Activity can be set for 1-24 hours in one-hour increments. Upon cancellation or
expiration of the dened time period, full automated insulin delivery starts on its own
and SmartAdjust™ technology returns to using the Target Glucose dened in settings.
Beginning Automated Mode
An active Pod and CGM transmitter serial number are required to switch to Automated Mode. Users can
switch to Automated Mode immediately with the rst Pod.
With the first Pod wear, since no recent insulin delivery history is available, SmartAdjust™
technology uses the user-entered Basal Program as a baseline to initiate automated insulin
delivery. As a safety measure, automated insulin delivery will be more restrained for the rst Pod.
With subsequent Pod changes, SmartAdjust™ technology uses the “adaptive basal rate” as a
baseline for automated insulin delivery.
- The adaptive basal rate is based on insulin delivery history.
- Adaptivity adjusts with each Pod change to meet individual insulin needs.
Omnipod® 5 System Start
Omnipod® 5 System Management and Optimization
Screen images shown in this guide are examples only and
are not suggestions for user settings.