For questions call: 1-800-634-4350
Pads are for Single Patient Use Only.
OmronHealthcare.com
3125420-0E

Questions and Answers

  • T
    Terry ScottAug 17, 2025
    Why is my Omron Avail PM601 Medical Equipment intensity too weak?
    • S
      Samantha CarrAug 17, 2025
      The intensity may be weak for several reasons: the pads may be stacked or overlapping (check their placement), TENS therapy might not be selected ('MICROCURRENT' will prevent you from feeling intensity), the intensity setting could be too low (increase it in the app), or the pad's adhesive surface may be damaged (replace the pad).
  • C
    Charles TorresAug 20, 2025
    What to do if my Omron Avail PM601 Medical Equipment power cuts off during use?
    • J
      judywarrenAug 20, 2025
      The power may be cutting off because the battery charge is extremely low (recharge it fully). Also, the device automatically turns off 10 minutes after therapy completion, pad removal, or therapy pausing, so check if any of these have occurred and ensure the pad is correctly placed on your body.
  • M
    Michael TravisAug 22, 2025
    Why won't my Omron Avail PM601 device recharge?
    • W
      wolfkellyAug 22, 2025
      The device might not be recharging because the battery charger isn't connected properly, so check the AC adapter connection to both the charger and the power outlet. Also, ensure the device is properly placed on the charger without any foreign objects. High heat or cold temperatures can also stop the battery from recharging, potentially increasing the charging time.
  • K
    Kimberly ParkAug 25, 2025
    What to do if Omron Avail PM601 shows connection failure?
    • B
      Bryan KruegerAug 26, 2025
      If your device shows a connection failure, it might be too far from your smartphone. Move the device within 16 ft (5 m) of the smartphone. Also, consider if the Omron TENS app was uninstalled and reinstalled.
  • S
    sarahjacksonAug 28, 2025
    How to fix Omron Avail PM601 Medical Equipment pad not sticking to skin?
    • L
      lisasmithAug 28, 2025
      If the pad isn't sticking, make sure you've removed the transparent film from the adhesive surface. Also, if the pad or your skin is wet, air dry both surfaces. If the adhesive surface is damaged, replace the pad. Shave the area if there is too much hair on your skin. Replace the pad if it was stored under high temperature, high humidity or direct sunshine.
  • T
    Timothy SchultzAug 30, 2025
    What to do if Omron Medical Equipment skin turns red or feels irritated?
    • F
      fishermelanieAug 30, 2025
      If your skin becomes red or irritated, shorten the therapy session to less than 30 minutes. Ensure the pads are placed correctly on your body, and replace the pad if the adhesive surface is worn out.
  • R
    Rachel BarnettSep 2, 2025
    How to fix Omron Avail PM601 Medical Equipment unable to tap start button?
    • H
      hunterkyleSep 2, 2025
      If you're unable to tap the 'START' button, check that the pads are properly placed on your body, and ensure the device is correctly attached to the pad. Also, verify that you've removed the transparent film from the pad's adhesive surface. The pad may also be damaged, so consider replacing it. Also, make sure that the device power is on and connected to the app.
  • M
    Margaret AlvaradoSep 5, 2025
    Why won't my Omron Avail PM601 device turn on?
    • K
      Katherine WolfSep 5, 2025
      The device might not be turning on because the batteries are depleted. If so, recharge the battery fully.
  • K
    Kenneth KimSep 7, 2025
    What does blinking purple light mean on Omron Avail PM601 Medical Equipment?
    • F
      fergusonjosephSep 8, 2025
      The blinking purple light indicates abnormalities in the device. Stop using it immediately.
  • R
    Rebecca NunezSep 12, 2025
    What to do if Omron Avail PM601 Medical Equipment devices are hot when charging?
    • J
      Joshua MannSep 12, 2025
      If the devices, charger, or AC adapter are abnormally hot during charging, they may be damaged. Unplug the AC adapter from the charger and remove the device from the charger immediately.

Summary

Introduction and Safety Guidelines

Safety Instructions

Important safety information for safe and proper use of the device.

Intended Use of the Device

Device's purpose for pain relief in muscles and joints due to strain.

Contraindications for Use

Reasons why the device should not be used, e.g., pacemakers.

Critical Safety Information

General Warnings

Critical safety precautions to prevent serious injury or death.

Usage Cautions

Precautions to prevent minor injury or equipment damage.

Understanding the Device Functionality

Device and App Overview

Operating Instructions

Getting Started with Your Therapy

Pad Placement Guidelines

Guidelines for cleaning the skin and correctly placing the pads on various body parts.

Areas to Avoid for Pad Placement

Critical areas where pads should not be applied due to safety risks.

Pain Management Strategies

Cleaning and Disposal

Appendix and Troubleshooting

Troubleshooting Common Issues

Common issues and their solutions for device operation and connectivity.

Device Specifications

Electromagnetic Compatibility (EMC)

Omron Avail PM601 Specifications

General IconGeneral
CategoryMedical Equipment
Device TypeBlood Pressure Monitor
ModelPM601
TechnologyOscillometric
Battery LifeApproximately 300 measurements (with new alkaline batteries)
Data Storage60 readings
Intended UseHome blood pressure monitoring
Cuff Size22-32cm
Pressure Range0-299mmHg
Pulse Rate Range40 to 180 beats/min