OMRON ULTRASONIC
NEBULISER
Model NE-U17
(NE-U17-E)
Instruction Manual
Thank you very much for purchasing this OMRON
Ultrasonic Nebuliser.
Be sure to read this instruction manual carefully
before using the unit so that you can use it safely and
correctly.
Please keep this instruction manual always at hand
for future reference.
Ultrasonic
Nebuliser
NE-U17
AIR-FLOW VOLUME
NEBULISATION VOLUME
TIMER(MIN)
WATER LEVEL
MEDICATION CUP
COVER
FAN COVER
TIMER
POWER
CONTINUOUS
CONTINUOUS
START/STOP
Table of Contents
Be sure to Read This Section
Exemptions from Liabilities ......... 1
Intended Use .............................. 2
Notes on Safety .......................... 3
About the Product
Features of the Product .............. 6
Composition of the Product ........ 7
Description of the LCD Display .. 7
Names of the Parts .................... 8
How to Use the Unit
How to Prepare the Unit
before Inhalation ...................... 9
How to Inhale ........................... 11
How to Inhale
(Using Optional Parts) ........... 13
How to Take Care of the Unit
How to Take Care of the Unit ... 15
How to Disinfect the Unit .......... 17
How to Store the Unit ............... 18
Specifications
Troubleshooting ........................ 19
Specifications ........................... 20
List of Optional Parts ................ 22
IM-U17-E-03-09/2014
1615886-5G
This unit is a medical instrument. Be sure to use the unit properly
according to instructions given by medical professionals.
Only use medications as prescribed and instructed by your doctor.
EN
DE
NL
FR
IT
ES

Questions and Answers

  • S
    Sara MartinezAug 2, 2025
    What to do if my Omron NE-U17 doesn't nebulize?
    • T
      Teresa MaciasAug 2, 2025
      If your Omron Medical Equipment isn't nebulizing, several factors could be at play. Ensure the water in the operating chamber is filled to the specified level. The medication cup must be set correctly, and the fan cover needs to be properly closed and the fan set properly. If there's too much medication (over 150 ml) in the cup, reduce the amount. Also, check the inhalation hose for bends and pooled medication; if found, adjust the hose and drain the medication.
  • K
    Kathy FieldsAug 10, 2025
    Why is the nebulization rate low on my Omron NE-U17?
    • H
      hansonerinAug 10, 2025
      A low nebulization rate on your Omron Medical Equipment could stem from several causes. If there's too little medication (less than 5 ml), increase the amount. If the medication cup is damaged, replace it. If there's too much medication in the medication cup, reduce it to less than 150 ml. Ensure the nebulization rate or air volume adjustment knob isn't set too low by turning the knob to the right to adjust the value. If the room or water temperature is too low, operate the unit without medication for about 4 minutes before use. Also, ensure the medication cup is clean and free of detergent residue.
  • M
    Michael McdonaldAug 17, 2025
    What to do if nebulization is unstable with my Omron NE-U17 Medical Equipment?
    • B
      Brett MacdonaldAug 17, 2025
      If you're experiencing unstable nebulization with your Omron Medical Equipment, it might be due to the amount of medication exceeding 150 ml; reduce it accordingly. You may also need to adjust the nebulization rate.
  • M
    munozjamesAug 25, 2025
    What to do if my Omron NE-U17 Medical Equipment displays '3 2 1 WATER LEVEL MEDICATION CUP COVER FAN COVER'?
    • C
      cheyenne81Aug 26, 2025
      This display on your Omron Medical Equipment indicates a few potential issues. Ensure the water level in the operating chamber is filled to the correct level. Check that the medication cup cover and the fan cover are securely closed. If the fan inside the main unit has stopped, contact the nearest OMRON dealer.

Summary

Exemptions from Liabilities

Intended Use

Safety Instructions and Precautions

General Usage Safety Advice

General precautions and advice for using the ultrasonic nebuliser safely and correctly.

Unit Maintenance and Safety

Instructions and confirmation steps before using the unit for operation.

Handling Operational Faults and Accidents

Measures to take when an equipment error (E1) occurs or a fault is detected.

Product Features Overview

Product Composition and Display Details

LCD Display Functionality

Explanation of the On-Screen Display elements and their meanings for the ultrasonic nebuliser.

Nebuliser Parts Identification

Unit Preparation for Inhalation

Inhalation Procedure Guide

Inhalation with Optional Parts

Nebuliser Maintenance and Care

Air Filter Replacement Guide

Procedure for replacing the air filter when it becomes dirty.

Fan and Cover Cleaning Instructions

Instructions for cleaning the fan and fan cover.

Main Unit Cleaning Procedure

Guidance on cleaning the main unit with water or alcohol.

Nebuliser Disinfection Methods

Disinfection and Storage Guidelines

Safe Unit Storage Practices

Guidelines for storing the parts after drying to prevent infection or damage.

Nebuliser Troubleshooting Guide

Technical Specifications

Optional Nebuliser Accessories

Omron NE-U17 Specifications

General IconGeneral
TypeUltrasonic Nebulizer
Product NameNE-U17
CategoryMedical Equipment
Operation ModeContinuous
Particle Size5 μm
AccessoriesMouthpiece

Related product manuals