9
How to Prepare the Unit before Inhalation
Clean the parts before starting the inhalation.
[Parts that can be disinfected by boiling]
(For the disinfecting method, refer to Page 17.)
[Parts that cannot be disinfected by boiling]
(For the care and disinfection, refer to Page 17.)
Submerge these parts in a container filled with
sufficient water and boil for 10 minutes.
Medication
cup cover
Washer
Be sure to immediately dry the cleaned and disinfected
parts, then store them in a dry place in order to prevent
them from getting contaminated in the future.
1
Filling the cooling water.
1) Remove the nebulisation section
by rotating the medication cup
cover-fixing lever clockwise.
2
1
water level
2) Fill the cooling water.
Fill the cooling water up to the “water
level” mark that specifies the water level
in the water tank. The proper amount is
approximately 375 ml.
water level
water level
If the water level in the water tank is lower than
the “water level” mark, a “lower water level in water
tank” error will be displayed.
If the water temperature in the water tank is low,
the nebulisation rate may drop. (Recommended
temperature is approximately 26°C (78.8°F).
Replace water in the water tank each time the unit
is used.
When the unit is used for a long time continuously,
replace water at least every twenty-four hours.
(For continuous use or long periods, refer to
Page 13.)
Mouthpiece
set
Air duct cover
Fan cover
Air filter case
Air filter
Fan
Medication cup
Medication cup holder
Warning
Inhalation hose (M)
with a cuff
How to remove the washer
Place the medication cup cover
on its side and insert a small item
(such as a paper clip) into the
hole of the cover.
Pinch and remove the washer
gently.
How to attach the washer
Hold the medication cup cover and
place the washer securely with
your fingertips as illustrated in the
figures on the left.
Make sure the washer is placed
flat and not upside down.