IMPORTANT INFORMATION REGARDING ELECTRO MAGNETIC COMPATIBILITY EMC
Electromagnetic immunity IEC60601-1-2
Immunity test IEC 60601 Test level Compliance level
Electromagnetic environment –guidance
Electrostatic discharge,
ESD
(IEC 61000-4-2)
Contact discharge:
±8 kV
Air discharge:
±15 kV
Contact discharge:
±8 kV
Air discharge:
±15 kV
The relative humidity should be at least 5 %.
Electrical fast transients /
bursts
(IEC 61000-4-4)
Power supply lines: ±2 kV
Longer input / output lines:
±1 kV
Power supply lines: ±2 kV
Longer input / output lines:
N/A
Mains power quality should be that of a typical home,
commercial or hospital environment.
Surges on AC mains lines
(IEC 61000-4-5)
Common mode:
±2 kV
Differential mode:
±1 kV
Common mode:
N/A
Differential mode:
±1 kV
Mains power quality should be that of a typical home,
commercial or hospital environment.