10. Specifications
Model
BP5255 HEM-716BT2-ZAZ
Display LCD digital display
Cuff pressure range 0 to 299 mmHg
Blood pressure SYS: 60 to 260 mmHg
measurement range DIA: 40 to 215 mmHg
Pulse measurement range 40 to 180 beats / min.
Pressure: ±3 mmHg
Accuracy
Pulse: ±5 % of display reading
Inflation Automatic by electric pump
Deflation Automatic pressure release valve
Measurement method Oscillometric method
Transmission method
Bluetooth
®
Low Energy
Frequency range: 2.4 GHz (2400 - 2483.5 MHz)
Wireless communication Modulation: GFSK
Effective radiated power: < 20 dBm
Operation mode Continuous operation
Monitor: IP20
IP classification
Optional AC adapter: IP21
Rating DC6 V 4 W
4 “AA” batteries 1.5 V or optional AC adapter (INPUT
Power source
AC100 - 240 V 50 - 60 Hz 0.12 - 0.065 A)
Approximately 500 measurements (using new alkaline
Battery life
batteries)
Monitor: 5 years or the time when reaching 10000
times of use.
Durable period (Service
Arm cuff: 5 years or the time when reaching 10000
life)
times of use.
Optional AC adapter: 5 years
35