Loading...
37
appears
The pulse rate is not
detected correctly.
Apply the arm cuff
correctly, then take
another measurement.
Refer to sub-section 2.4.
Remain still and sit
correctly during a
measurement.
If the “ ” symbol
continues to appear, we
recommend you to
consult with your
physician.
/
appears
does not flash during
a measurement
appears
The monitor has
malfunctioned.
Press the [START/
STOP] button again. If
“Er” still appears, contact
customer service. Refer
to section 12.
appears
The monitor cannot
connect to a smart
device or transmit data
correctly.
Follow the instructions
shown in the app. If the
“Err” symbol still appears
after checking the app,
contact customer
service. Refer to
section 12.
Display/Problem Possible Cause Solution