7. Error Messages and Troubleshooting
EN
Movement during measurement and the
arm cuff has not been inflated sufficiently.
Repeat measurement. Remain still and do not
talk during measurement.
Refer to chapter 3.
If “E2” appears repeatedly, inflate the arm cuff
manually until it is 30 to 40 mmHg above your
previous measurement result.
Refer to chapter 3.
The arm cuff was inflated exceeding the
maximum allowable pressure, and then
deflated automatically.
Do not touch the arm cuff and/or bend the air
tube while taking a measurement. Do not
inflate the arm cuff more than necessary.
Refer to chapter 3.
Movement during a measurement.
Repeat measurement. Remain still and do not
talk during measurement.
Refer to chapter 3.
Movement during a measurement.
Clothing is interfering with the arm cuff.
Remove any clothing interfering with the arm
cuff.
Refer to section 2.4.
Connection failure. /
Data is not being transmitted.
Refer to “Connection failure. / Data is not
being transmitted.” in section 7.2.
Device error. Contact Customer Service.
Display Possible Cause Solution

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron BP769CAN and is the answer not in the manual?

Omron BP769CAN Specifications

General IconGeneral
BrandOmron
ModelBP769CAN
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals