106
3-8. Advanced Registration of Patient Information
Registration Procedure
Press the [MAIN MENU] button in the initial screen. Perform "Register / Manage patient info" from
the main menu screen.
A. Importing a text file
1. You will need a USB flash drive.
2. Create the data on a computer.
Create the data as follows:
Data format: text file*
1
File name: PATIENT.CSV
ID: Alphanumeric characters (upper case), hyphens (-); maximum of 13 digits
Name: Alphabet, maximum of 40 characters
Sex: Code input; female = 0, male = 1
Birth date: Numeric characters, / (slashes); yyyy/mm/dd (zeros can be omitted)
Height: Numeric characters, 120 to 210, units = cm
Weight: Numeric characters, 25.0 to 300.0, units = kg
Waist: Numeric characters, 30 to 250, units = cm
Disease: Code input*
2
, 1 character
Measurement site: Code input*
2
, 1 character
Doctor: Character type same as for "Name"
Technician: Character type same as for "Name"
Category: Character type same as for "Name"
Order number: Numeric characters, hyphens(-); maximum of 16 digits
*
1
An XML file can also be imported. To use an XML file, consult
a dealer or an Omron Healthcare technical support
representative.
*
2
Code Table (refer to page 107)
3. Save the created data to USB flash drive and connect it to the USB port on the device.
4. Press the [REGISTER / MANAGE PATIENT INFO]
button.
5. Press the [ADD NEW ID] button.
ABC,
DEF,
GHI,
JKL,
Internal,
Internal,
BP-203RPE3(A).fm 106 ージ 󲥿󲥽󲥽9年󲥾󲥿月󲥿6日 土曜 午5時󲥿分