30
Other problems
Error
code
Description Points to check
E3
Cuff inÀ ated to 300 mmHg or
higher during manual inÀ ation
in auscultation mode
When inÀ ating manually in
auscultation mode, release
the button when the pressure
reaches the desired value.
Over-inÀ ation occurs
If this occurs during
measurement, repeat
measurement.If this occurs
when not performing
measurement, contact your
local OMRON representative.
E40 The batteries are depleted.
Replace with new batteries.
(page 13)
E42 Voltage error
Replace the batteries with
new batteries or new AC
adapter. If the error continues,
contact your local OMRON
representative.