Recommended Separation Distances:
Recommended separation distance between portable and mobile RF
communications equipment and the HBP-1100
The HBP-1100 is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
HBP-1100 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the HBP-1100 as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power of
transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1,2
P
80 MHz to 800 MHz
d = 1,2
P
800 MHz to 2.5 GHz
d = 2,3
P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
11,21,22,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance
d
in metres (m) can be determined using the
equation applicable to the frequency of the transmitter, where
P
is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note1:At 80MHz and 800MHz, the separation distance for the higher frequency
range applies
Note2:These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reÀ ection from structures,
objects, and people.