35
Manufacture's Declaration
Immunity 
Test
IEC60601-1-2
Test Level
Compliance 
Level
Electromagnetic Environment 
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V r.m.s
150 kHz ~ 80 MHz
80% AM (2Hz)
3 V/m
80 MHz ~ 800 MHz
3 V/m
800 MHz ~ 2.5 GHz
3 V r.m.s
3 V/m
3 V/m
Potable and mobile RF communications 
equipment should be used no closer to 
any part of the HBP-9020 including 
cables, than the recommended separation 
distance calculated from the equation 
appropriate to the frequency of the 
transmitter.
Recommend separation distance
d = 1.2   150 KHz to 80 MHz
d = 1.2   80 MHz to 800 MHz
d = 2.3   800 MHz to 2.5 GHz
where P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter manufacturer 
and d is the recommended separation 
distance in meters (m).
Field strengths from fixed RF transmitters 
as determined by an electromagnetic site 
survey* should be less than the 
compliance level in each frequency 
range.** 
Interference may occur in the vicinity of 
equipment marked with the following 
symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects, and people.
* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones 
and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an 
electromagnetic site survey should be considered. If the measured field strength in the location in which 
the HBP-9020 is used exceeds the applicable RF compliance level above, the HBP-9020 should be 
observed to verify normal operation. If abnormal performance is observed, additional measures may be 
necessary, such as reorienting or relocating the HBP-9020.
** Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
HBP-9020.fm  35 ページ  2010年7月9日 金曜日 午前8時59分