9
Names and Functions of the Parts
Functions and setting while the unit is in use
(1) Pressure setting function
A function that can switch the pressure setting
Inflation can be set by AUTO (automatic setting) method or the pressure value setting method.
AUTO (Automatic setting): Can be used when the SINGLE, AVG., or MANU. Mode is selected.
The unit estimates the systolic blood pressure value during inflation and inflates to an appropriate
cuff pressure.
Pressure value setting method: Pressure value can be set between 100 and 280 mmHg. Set the
pressure value at about 30 to 40 mmHg higher than the expected systolic pressure.
• To set the P-SET to "AUTO," turn the P-SET Knob counterclockwise as far as it goes until you can hear a click sound.
• In the "AUTO" setting, proper inflation may not be possible when the systolic blood pressure is expected to be more than 220
mmHg. Use the unit in the pressure value setting method.
• If the cuff cannot be inflated to the pressure necessary for the measurement, it may be re-inflated automatically.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P
-
SET
100
AUTO
140 180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P
-
SET
100
AUTO
140 180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
(2) Non-display function
A function not to display measurement results. However, the cuff
pressure during measurement is displayed.
This function can be used in the SINGLE and AVG. Modes.
By each push of the Button, display or non-display of
the measurement results is switched.
HIDE
(3) Manual deflation control function
A function to accelerate deflation by pushing the DEFLATION
(deflation control) / Display Switch Button during deflation.
This function can be used in the MANU. Mode.
With each push of the Button, deflation speed is
accelerated by 5 to 10 mmHg.
BPM
mmHg
PULSE
SYS
mmHg
DIA
AC
AC
Display Non-display
P
-
SET
HIDE
HEM-907-draft(Eng) 9/21/00 4:07 PM Page 9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron HEM-907 and is the answer not in the manual?

Omron HEM-907 Specifications

General IconGeneral
ModelHEM-907
Measurement MethodOscillometric
Power SourceAC Adapter or 4 AA Batteries
Pressure Range0 to 299 mmHg
Accuracy Pressure±3 mmHg
Accuracy Pulse±5% of reading
InflationAutomatic by electric pump
DeflationAutomatic pressure release valve
Irregular Heartbeat DetectionYes
Movement DetectionYes
TypeDigital Blood Pressure Monitor
DisplayLCD
Pulse Range30 to 199 beats/min
Cuff Size22-32 cm (standard), 32-42 cm (large)