K3NX K3NX
4
3
MEASURING RANGES
Input range Measuring range Input impedance Reliability (See Note 2.) Instantaneous overload
(30 seconds)
DC voltage
æ
±
199.99 V
10 M
Ω
±
0.1%rdg
±
1 digit max.
±
400 V
å
±
19.999 V
1M
Ω
±
200 V
ä
±
1.9999 V
10 M
Ω
min.
ã
±
199.99 mV
10 M
Ω
min.
â
1.0000 to 5.0000 V 1M
Ω
DC current
æ
±
199.99 mA
1
Ω
±
0.1%rdg
±
1 digit max.
±
400 mA
å
±
199.99 mA
10
Ω
±
200 mA
ä
±
1.9999 mA
100
Ω
ã
4.000 to 20.000 mA 10
Ω
AC voltage
æ
0.0 to 400.0 V 1M
Ω
±
0.3%rdg
±
5 digit max.
700 V
å
0.00 to 199.99 V 1M
Ω
ä
0.000 to 19.999 V 1M
Ω
±
0.5%rdg
±
10 digit max.
400 V
ã
0.0000 to 1.9999 V 10 M
Ω
min.
AC current
æ
0.000 to 10.000 A (0.5 VA CT)
(See Note 4.)
±
0.5%rdg
±
20 digit max.
20 A
å
0.0000 to 1.9999 A (0.5 VA CT)
(See Note 4.)
ä
0.00 to 199.99 mA 1
Ω
±
0.5%rdg
±
10 digit max.
2A
ã
0.000 to 19.999 mA 10
Ω
Note: 1. The “rdg” stands for “reading value.”
2. The accuracy is guaranteed for the input frequency range of 40 Hz to 1 kHz (except for
æ
and
å
ranges of AC current input) and
the ambient temperature of 23
±
5
°
C.
If the actual input in each of the following measuring ranges is 10% of the maximum value or less, the following accuracy values
will apply.
Input range
Reliability
DC voltage
æ
,
å
,
ä
,
ã
,
â
±
0.15% FS
DC current
æ
,
å
,
ä
,
ã
±
0.1% FS
AC voltage
æ
±
0.25% FS
å
±
0.5% FS
ä
,
ã
±
0.15% FS
AC current
æ
±
0.15% FS
å
±
0.1% FS
ä
,
ã
±
1.0% FS
3. When using a DC voltage input model in the
ä
and
ã
range, do not open the input terminals. The input terminals can be opened,
however, if a resistor of approximately 1 M
Ω
is connected to the input terminals.
1234
±
1.9999 V
±
199.99 mV
+--
1M
Ω
4. “0.5 VA CT” indicates consumption VA of the internal CT.