OMS ARCADIA EXT S.T.01/3R ED.0 REV.8 09/2020
53
6.4.2. REPLACEMENT OF FUSES
ATTENTION: In order to replace a fuse, you must first identify the causes that have caused the rupture and only
after removing them proceed with the replacement. The new protection fuse must have the same
characteristics as the one replaced, indicated on the electrical diagram of the device.
On the base of the armchair, the protection fuses (B) and (C) can easily be identified, that can be replaced
externally by means of a screwdriver
A series of spare fuses provided by O.M.S. Is located inside the accessory box
You can see here below the values of the main fuses, for the sole reference only:
A. General Power of the Armchair 6.3AT, 230 Vac
B. Transformer power supply 4.0 AT, 230 Vac
DANGER
The main switch realizes the insulation of the device from the direct power supply,
before making any operation inside the device make sure that switch is off.