Appendix A: Electromagnetic Compatibility (EMC) Information
AutoVector
®
Fluid Immersion Simulator
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
The use of Accessories, transducers, and cables other than those specied by the manufacturer, may result in
increased Emissions or decreased Immunity of the Auto Vector Fluid Immersion Simulator.
The AutoVector Fluid Immersion Simulator should not be used adjacent to or stacked with other equipment and
that if adjacent or stacked use is necessary, the Auto Vector Fluid Immersion Simulator should be observed to
verify normal operation in the conguration in which it will be used.
Guidance and Manufacturer’s Declaration –Electromagnetic Emissions
The AutoVector Fluid Immersion Simulator is intended for use in the electromagnetic environment specied
below. The customer or the user of the Auto Vector Fluid Immersion Simulator should assure that it is used in
such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The AutoVector Fluid Immersion Simulator uses RF energy only for
its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The AutoVector Fluid Immersion Simulator is suitable for use in all
establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations /
icker emissions
IEC 61000-3-3
Complies