20
CONTROL SOLUTION TESTING
Control solution testing
When to test with control solution
You may check your meter and test strips using the OneTouch® Select™ Control Solution.
OneTouch® Select™ Control Solution contains a known amount of glucose and is used to check that the meter and the test
strips are working properly. Remember which control solution you used.
CAUTION: Do Not swallow control solution; it is not for human consumption. Do Not apply control solution to the
skin or eyes as it may cause irritation.
Do a control solution test:
– whenever you open a new vial of test strips
– if you suspect the meter or test strips are not working
properly
– if you have had repeated unexpected blood glucose
results
– if you drop or damage the meter
NOTE:
• Use only OneTouch® Select™ Control Solution with your
OneTouch® Select™ Meter.
• Control solution tests must be done at room temperature
(20–25°C). Make sure your meter, test strips, and control
solution are at room temperature before testing.
• Write the discard date (3 months after rst opening the
vial) on the vial when you rst open it.
How to test with control solution
Start with the meter o. If you have turned the meter on to change settings or review past results, turn it o.
Check the code on the test strip vial before inserting the test strip
Insert a test strip to turn on the meter
Make sure the three contact bars are facing you and the arrow is pointing down. Push the test
strip in as far as it will go. Do Not bend the test strip.
Match the code on the meter with the code on the test strip vial
If the code on the meter does not match the code on the test strip vial, press
or
to
match the code number on the test strip vial. The new code number will ash on the display
for three seconds, then briey stop ashing, after which the display will advance to the
APPLY SAMPLE screen.
If the codes already match, press
to go to the APPLY SAMPLE screen. When you do not make
a change after ve seconds, the display will advance to the APPLY SAMPLE screen.
Mark the test as a control solution test
Press
to change APPLY SAMPLE to APPLY CONTROL. You must mark the test before you
apply control solution. Once you have completed the test, you cannot change the marking.
The meter is now ready to perform a control solution test.
APPLY CONTROL
▲