Electrical and safety standards
This meter complies with CISPR 11: 2003, Class B (Radiated Only). Emissions of the energy used are low and not likely to cause interference
in nearby electronic equipment. The meter has been tested for immunity to Level 3 electrostatic discharge as specied in IEC 61000-4-2.
This meter has been tested for immunity to radio frequency interference over the frequency range 80 MHz to 2.5 GHz at 3 V/m as specied
in IEC 61000-4-3.
Degree of protection rating: IP31
To maximize your chances of an accurate comparison between meter and laboratory results, follow a few basic guidelines:
Before going to the lab
• Perform a control solution test to make sure the meter is working properly.
• Do Not eat for at least eight hours before you test your blood.
• Take your meter with you to the lab.
While at the lab
• Conduct your meter test within 15 minutes of the lab test.
• Use only fresh, capillary blood obtained from the ngertip.
• Follow all instructions in this User Guide for performing a blood glucose test with your meter.
Reported result range 20–600 mg/dL
Calibration Plasma-equivalent
Sample Fresh capillary whole blood
Test time 5 seconds
Assay method Glucose oxidase biosensor
Meter power source One replaceable 3.0 Volt CR 2032
lithium battery (or equivalent)
Unit of measure mg/dL
Memory 350 blood glucose or control
solution test results
Automatic shuto 2 minutes after last action
Size 90 mm x 55.54 mm x 21.7 mm
Weight Approximately 52.21 g,
with battery
Operating ranges Temperature: 10–44°C
Relative humidity: 10–90%
Altitude: up to 3048 meters
Hematocrit: 30–55%
Battery ratings 1 x 3.0 V d.c., 60 mA
(1 x CR 2032 batteries)
direct current
Technical specications
Symbols
Cautions and Warnings. Refer to safety-related notes in this User Guide and inserts that came with your meter and testing supplies.
Low battery
Direct Current
Manufacturer
Serial number symbol
Lot Number
2
Do not re-use
In Vitro Diagnostic Medical Device
Expiration Date
Sterilized by irradiation
Consult Instructions for Use
Contains sucient for <n> tests
Storage temperature limits
Recycle
Guarantee
LifeScan guarantees that the OneTouch® Select™ Meter will be free of defects in material and workmanship for three years, valid from the
date of purchase. The guarantee extends only to the original purchaser and is not transferable.
The system described herein is covered by one or more of the following U.S. patents: 5,708,247, 5,951,836, 6,241,862, 6,284,125, and
7,112,265. Use of the monitoring device included herein is protected under one or more of the following U.S. patents: 6,413,410, 6,733,655,
7,250,105, 7,468,125. Purchase of this device does not act to grant a use license under these patents. Such a license is granted only when the
device is used with OneTouch® Select™ Test Strips. No test strip supplier other than LifeScan is authorized to grant such a license. The accuracy
of results generated with LifeScan meters using test strips manufactured by anyone other than LifeScan has not been evaluated by LifeScan.