39
Troubleshooting
Your control
solution test
result is below
1.1 mmol/L.
Test again with control
solution and a new test
strip. If the result is LOW
GLUCOSE
again, do not
use the meter. Contact
Customer Care on
0800 121 200 (UK) or
1800 535 676 (Ireland).
Message What it means What to do
You may have
a very high
blood glucose
level (severe
hyperglycaemia),
above
33.3 mmol/L.
Re-test your
blood glucose. If
the result is HIGH
GLUCOSE
again,
consult your health
care professional
and follow his/her
instructions
without delay.
Your control
solution test
result is above
33.3 mmol/L.
Test again with control
solution and a new test
strip. If the result is HIGH
GLUCOSE
again, do not
use the meter. Contact
Customer Care on
0800 121 200 (UK) or
1800 535 676 (Ireland).
WARNING
AFTER MEAL?
HIGH GLUCOSE
ABOVE 33.3 mmol/L
WARNING
CONTROL TEST
LOW GLUCOSE
BELOW 1.1 mmol/L
WARNING
CONTROL TEST
HIGH GLUCOSE
ABOVE 33.3 mmol/L