Troubleshooting
Message What it means What to do
Meter is too
cold (below
10°C) to work
correctly.
WARNING
SEE USER’S HANDBOOK
TEMPERAT URE
ERROR
TEMPERATURE
TOO LOW
Meter is too
hot (above
44°C) to work
correctly.
WARNING
SEE USER’S HANDBOOK
TEMPERAT URE
ERROR
TEMPERATURE
TOO HIGH
Move the meter and
test strips to a cooler
area. Insert a new test
strip when the meter
and test strips are
within the operating
temperature range
(10–44°C). If you
do not get another
TEMPERATURE ERROR
message, you can
proceed with testing.
Move the meter and
test strips to a warmer
area. Insert a new test
strip when the meter
and test strips are
within the operating
temperature range
(10–44°C). If you
do not get another
TEMPERATURE ERROR
message, you can
proceed with testing.