Laser Power / Energy Meter User Manual

Table of Contents

Questions and Answers

  • E
    Emily MooreAug 17, 2025
    Why is my OPHIR StarBright reading incorrectly?
    • A
      Amanda HoustonAug 17, 2025
      If your OPHIR Measuring Instruments device shows incorrect or erratic readings, especially on a sensitive scale, it might be due to electromagnetic interference from the pulsing laser causing misreading or false triggering.
  • C
    chenkristopherAug 19, 2025
    What to do if OPHIR StarBright triggers without laser pulses?
    • D
      dballAug 19, 2025
      If your OPHIR Measuring Instruments device triggers without laser pulses, ensure the sensor is mounted to the stand using the insulated plastic rod provided, and not a metal rod.
  • M
    Mrs. Tammy KellerAug 22, 2025
    How to troubleshoot OPHIR StarBright that won't operate after being discharged?
    • M
      mary72Aug 22, 2025
      If your OPHIR Measuring Instruments device doesn't operate after being completely discharged and connected to the charger, verify that the charger is providing 12V.
  • J
    Julia NormanAug 25, 2025
    Why does OPHIR StarBright Measuring Instruments work with charger but not battery?
    • C
      Christy SimpsonAug 26, 2025
      If your OPHIR Measuring Instruments device operates with the charger but not with the battery alone, recharge it overnight with the StarBright turned off for 5 hours. If it still does not work with the battery, the Lithium Ion battery is probably dead and needs replacement.
  • S
    sydney77Aug 28, 2025
    How to fix OPHIR Measuring Instruments that doesn't return to zero?
    • A
      Anthony PetersAug 28, 2025
      If your OPHIR Measuring Instruments device does not return completely to zero on power measurement: * If the sensor is very hot, allow it to cool. * Disconnect the sensor from the instrument. If the readout unit still doesn't zero, follow the instructions for Zero Adjustments. If the offset persists, try zeroing with the sensor connected, as described in the same section.
  • K
    Kyle GeorgeAug 30, 2025
    What to do if OPHIR Measuring Instruments triggers on noise?
    • T
      Tina TownsendAug 30, 2025
      If your OPHIR Measuring Instruments device triggers on background noise or sometimes fails to catch a large pulse, increase the threshold level. See Set Threshold.
  • H
    Heather BlackSep 2, 2025
    Why does OPHIR StarBright show a high frequency?
    • C
      Colton MyersSep 2, 2025
      If your OPHIR Measuring Instruments device shows a frequency that is too high: 1. Try keeping the cable away from the bench. 2. Move the sensor/display further away from EMI.
  • M
    Monique KeithSep 5, 2025
    What to do if OPHIR StarBright Measuring Instruments is not ready after reading?
    • C
      Colleen JacksonSep 5, 2025
      If your OPHIR Measuring Instruments device does not show ready for a long while after a reading is made, increase the threshold level. See Set Threshold.
  • V
    Veronica DudleySep 8, 2025
    Why does OPHIR Measuring Instruments show zero reading?
    • C
      Christopher CooperSep 8, 2025
      If your OPHIR Measuring Instruments device shows a zero reading in both power and energy modes: * Check the connections between the sensor and the instrument (see Maintenance of Thermopile Sensors). * Check that the sensor disc is operative. Resistance between pins 1 and 9 of the sensor connector should be about 1.8k. If the sensor is defective, there will be an open or short circuit.
  • M
    Maria BushSep 10, 2025
    How to fix OPHIR StarBright Measuring Instruments with non-reproducible results?
    • J
      Joseph WilsonSep 12, 2025
      If you are getting non-reproducible results when measuring very small energy pulses, or no response to pulses at low energy with your OPHIR Measuring Instruments device, decrease the threshold level. See Set Threshold.

Summary

2 Quick Reference

2.1 Getting Started

Guides users on connecting sensors, powering on, and basic instrument settings.

2.2 Thermopile Sensors

Explains how to use StarBright with Thermopile sensors for power and energy measurements.

2.3 Photodiode Sensors

Details using StarBright with Photodiode sensors for power measurements and setting parameters.

2.4 Pyroelectric or Photodiode Energy Sensors

Covers using StarBright with energy sensors, including zeroing and parameter setup.

5 Thermopile Sensors

5.1 Operation of StarBright with Thermopile Absorber Sensors

Details connecting and operating StarBright with Thermopile absorber sensors.

5.2 Measuring Laser Power with Thermopile Sensors

Explains the procedure for measuring laser power using Thermopile sensors.

5.3 Measuring Laser Energy with Thermopile Sensors

Covers measuring single shot energy with Thermopile sensors.

6 BeamTrack Sensors

6.1 Setting Measurement Parameters

Guides on configuring parameters for BeamTrack sensors in Track mode.

6.2 Tracking Laser Beam Position and Size

Explains how to track beam position and size using BeamTrack sensors.

7 Photodiode Sensors

7.1 Operation of StarBright with Photodiode Sensors

Covers connecting and operating StarBright with Photodiode sensors.

7.2 Measuring Laser Power with Photodiode Sensors

Explains how to measure laser power using Photodiode sensors.

8 Pyroelectric and Photodiode Energy Sensors

8.1 Pyroelectric Sensors - Method of Operation

Explains the operational principles of Pyroelectric sensors.

8.2 Photodiode Energy Sensors - Method of Operation

Explains the operational principles of Photodiode energy sensors.

8.4 Operation of StarBright with Pyroelectric and Photodiode Energy Sensors

Guides on using StarBright with energy sensors, including zeroing.

9 Graphical Displays

10 Math Functions

11 Logging Data to a File for Upload to PC

12 Special Sensors

14 Calibration, Maintenance, and Troubleshooting

14.1 Updating Sensor Calibration Factors

Outlines general steps for updating sensor calibration factors for better accuracy.

14.2 Calibration of Thermopile Sensors

Covers absorber types and calibration methods for Thermopile sensors.

14.3 Calibration of Photodiode Sensors

Explains factory calibration and accuracy for Photodiode sensors.

14.4 Calibration of Pyroelectric Sensors

Details absorber types and calibration for Pyroelectric sensors.