optris PI – E2010-03-A
64
Analog output:
For voltage measurements the minimum load impedance should be 10KOhm.
The analog output can be used as a digital output. The voltage for “no alarm” and “alarm on” can be set within
the software.
The analog output (0 … 10V) has a 100 Ohm resistor in raw. With a maximum current of 10ma the voltage drop
is 1V.
Having an alarm LED with a forward voltage of 2V the analog output value for “alarm on” should be 3V as
maximum